Alignment Procedure for Attenuation Tester
Brace 2 ft. pipe into position. Disconnect monochromator. Put on pinhole.
Align laser to go through pinhole and give a spot on the center of the far box hole.
Position iris after stand for first lens. Close to a small hole and center on laser beam.
Place 50 cm fl lens 1 3/4" from front wall. Steer beam to be centered on far end.
[At this point it is possible to place tubes into beam and see how much steering the tubes do.]
Place 45
o
mirror and steer beam to center of second hole.
Place lens 5 1/2" from far wall. Steer beam to center of hole.
Close iris to 1" diameter.
Check vertical aperature on exit of monochromator. Vertical slits should be centered and opening should be 1/4".
Reconnect monochromator.