Summary of June 2000 CERN ISR Test Data

 

Updated : 19 July 2000 RL

 

    The following table summarizes the conditions of the data taken for the June 2000 CERN ISR Test.  It is primarily based on notations made in the FNAL logbook, but also makes use of the output log files, Event and Begin Run records, and headers of the datafiles.  You can find a sequential listing of all datafile headings and Begin Run records here. It is highly recommended you examine the Begin Run Records and Event Records of the data sets for a better understanding of the DAQ settings during a particular run.  A page to help you decode the Begin Run and Event Records are here.

 

You can find the June 2000 ISR Test Data Description here.

 

Notes:

 

Runs which have been checked and/or corrected are marked in YELLOW.

 

 

 

 

Initial System Setup and Diagnostics

RData 210

LDM Power off, 6 sensors

RData 211

LDM Power on, 6 sensors

RData 212

error w/ DCOPS (2 trys)

RData 323

new daq configuration, failed self test

RData324

corrupt data w/ 14 sensors

RData325

7 sensors, missing #0001 characters and sensor 1 data seems high

RData 326

6 sensors, run looks ok

RData 327

10 sensors, readout is ok,  10 cycles w/ 1 event per cycle (5 sec delay)

 

HP Unit Disconnected – Manual LDM Operation

RData 328

First trial w/ no Data

Second Trial w/ HP Disconnected, manual LDM operation (left on)

10 sensors connected

RData 329

Test w/ HP Disconnected

RData 330

Testing 13 sensor readout

1 event, data looks like empty buffers (no bkground)

RData 331

Testing 13 sensor readout

4 events, data looks like empty buffers

RData 332

Testing 13 sensor readout

20 events, data looks like empty buffers

RData 333

Test w/ HP Disconnected

RData 334

Test w/ HP Disconnected

RData 335

Test w/ HP Disconnected

RData 336

Test w/ HP Disconnected

RData 337

Test w/ HP Disconnected

RData 338

Test w/ HP Disconnected

RData 339

Axial Line Readout – No HP Data

RData 340

Axial Line Readout  w/ HP Data

RData 341

et0 readout only – looks like bkg only(?)

RData 342

et0 readout only, 1 event

 

Axial Line Surveyed, es7 micrometer moved from 2.90 -> 2.00

RData 343

20 cycles full Axial Line, No HP Data or Lasers On Data, abnormal termination after 16 cycles

RData 344

Aborted as test

 

Axial Line Complete - Axial Line Data Only w/ HP Unit Operational

RData 345

20 clusters, 1 event, 20 sec delay, 5 sec LDM warmup

RData 346

20 clusters, 1 event, 20 sec delay, 5 sec LDM warmup (small software correction)

RData 347

1 clusters, 20 events 5 sec LDM warmup

RData 348

Single Shot Axial Line

RData 349

Single Shot Axial Line

RData 350

Single Shot Axial Line

RData 351

Single Shot Axial Line

RData 352

Single Shot Axial Line

RData 353

Axial LDM has been cold 3 hours, doghouses put on et 3&4, Lights off,

20 cycles of quick data on Axial Line only

 

SLM Line Setup and Drilled Down

RData 354

Full SLM line – 1 shot 13 sensors

board reset in first event

RData 355

Full SLM line + Axial line

10 clusters, 4 events per cluster, 60 sec interval

abnormal termination, board reset in first event

RData 356

Full SLM line + Axial line

2 x 1 shot Readout w/ Wux and CC

Failed

 

Splitting Lines into SLM and Axial Lines 

Single Directional only unless noted

RData 357

SLM  Laser 303 – 5 events

RData 358

2 sensor stability s6 and s1 (es10 & es5)

20 Quick Measurements to compare peak quality and resolution of a ‘poor’ sensor (s1) and a ‘good’ one (s6)

No end run record, but data looks like it acquired ok

RData 359

2 sensor stability s6 and s1 (es10 & es5)

10 Quick Measurements to compare peak quality and resolution of a ‘poor’ sensor (s1) and a ‘good’ one (s6)

RData 360

2 sensor stability s6 and s1 (es10 & es5)

10 Quick Measurements to compare peak quality and resolution of a ‘poor’ sensor (s1) and a ‘good’ one (s6)

RData 361

Full SLM w/ 5 events

Abnormal termination in event 1

RData 362

Single Shot SLM

8 sensors readout

RData 363

Single Shot SLM

8 sensors readout

RData 364

2 Clusters SLM

1 Shot each

RData 365

SLM Line

60 Clusters, 1 event per cluster, 30 sec LDM warmup,

No end run record but readout of 60 clusters seems complete

At least through #0002 of last event in last cluster

RData 366

Aborted

RData 370

Axial Line

Single Shot Background Only (LDM was unplugged)

5 Sensors

RData 371

Axial Line

13 sensors connected, 5 readout

Cold laser, lights on, middle of afternoon

RData 372

Axial Line ( 5 sensors connected, 5 readout only)

25 cycles looking at LDM warmup again - Lights on, Doghouses on

Data initially has extra lines from DSP responses, but it appears limited to first event

RData 373

Axial Line ( 5 sensors connected, 5 readout only)

20 cycles looking at LDM warmup again - Lights on, Doghouses on

Data looks ok

 

Final SLM Tape Configuration Applied

RData 374

Axial Line

10 clusters w/ 3 events per cluster

Computer hung after completion – Data looks ok on simple inspection

RData 375

Axial Line running overnight

20 clusters w/ 4 events per cluster

60 sec spacing, 30 sec ldm warmup

Run aborted early upon return the next morning.

15/20 Clusters completed

 

Back to Single DCOPS String of 13 Sensors for SLM + Axial Lines

RData 376

13 sensors Bi-Directional

Laser 301 tied to Laser 303

4 Clusters 10:1 Laser ON:Laser OFF events

540 sec between clusters, 30 sec LDM warmup

84 events read

 

Laser 301 now tied to Laser 303 AND Laser 302 Permanently

RData 377

13 sensors Bi-Directional

Laser 301 tied to Laser 303 AND Laser 302

Single Shot

RData 378

13 sensors Bi-Directional

2 Clusters w/ 2 events each

No time delay between clusters

RData 379

13 sensors Bi-Directional

Long Overnight Run

70 Clusters 5:1 Laser ON:Laser OFF events

150 sec between  clusters, 30 sec LDM delay

Killed next morning before lights turned on

38.5 clusters ok – Lights on at cluster 34; cluster 22 & 32 not corrected

RData 380

Short Axial Run

20 clusters, 5 sensors only

60 sec delay, 30 sec LDM warmup

RData 381

Short Test Run

SLM + Axial Bi Directional

1 cluster w/ 1:1 laser on:laser off

RData 382

13 sensors Bi-Directional

2 Clusters, 3:1 Laser ON:Laser OFF events

90 sec delay, 30 sec LDM Warmup

 

Sensors Are Now Being Moved (RData 382 is original position)

RData 383

et2 moved +1mm in +X direction  (New position is 6mm)

Bi-Dir 2 Clusters w/ 4 Laser On per cluster

90 sec delay, 30 sec LDM warmup

Only 8 events taken (Single Shot Mode by mistake)

RData 384

Repeating  383

Bi-Dir 2 Clusters w/ 4 Laser On per cluster

90 sec delay, 30 sec LDM warmup

16 events taken

RData 385

et2 moved +1mm in +X direction  (New position is 7mm)

Bi-Dir 2 Clusters w/ 4 Laser On per cluster

90 sec delay, 30 sec LDM warmup

Z Sensor on MAB is lost now from change in position

System hung after 3 events

RData 386

Repeating 385

Bi-Dir 2 Clusters w/ 4 Laser On per cluster

90 sec delay, 30 sec LDM warmup

Z Sensor on MAB is lost now from change in position

System hung after 3 events

RData 387

et2 moved back to original position

Bi-Dir 2 Clusters w/ 4 Laser On per cluster

90 sec delay, 30 sec LDM warmup

RData 388

et2 moved -1mm in X direction 

Bi-Dir 2 Clusters w/ 4 Laser On per cluster

90 sec delay, 30 sec LDM warmup

 

Motion Shift Studies on SLM Line

et2 moved back to 0

All Data taken with 2 Clusters, 4 events per Cluster w/ Laser ON

 90 sec delay, 30 sec LDM warmup

RData 389

es5 moved 100 um in +Z (4 notches)

Previous position: .050

New Position: .054

RData 390

es5 moved 100 um in +Z (New Position: .058)

RData 391

es5 moved 100 um in +Z

RData 392

Back to .054

RData 393

Back to Original position (.050)

RData 394

es5 moved 100 um in -Z (New Position: .046)

1 event only

RData 395

Retry 394 w/ proper number clusters and events (described above)

RData 396

Original micrometer position

Aborted after 4 events by user (1 shot was intended)

RData 397

Repeat 396 but w/ One Shot Only

RData 398

Reading S5 only

Moving 20 notches tp .5mm

RData 399

Reading S5 only  - One Shot at .055

RData 400

Reading S5 only  - One Shot at .045

 

Ending Shift Studies

RData 401

Long Run

18 Clusters 4 Events per Cluster

90 sec delay, 30 sec LDM warmup

Lights Off w/ Doghouses on et 3&4

72 events read, but no end run record

 

Looking at Shifts once Again

RData 402

1 Shot of es5 (original position)

Lights On / Dog Houses On

** We have had problems w/ micrometer (see log book pg 73)

      Micrometer original position is .00505

RData 403

1 Shot of es5

3 sec LDM warmup

RData 404

1 Shot of es5

30 sec LDM warmup

RData 405

1 Shot of es5

30 sec LDM warmup

RData 406

1 Shot of es5

30 sec LDM warmup

RData 407

1 Shot of es5

30 sec LDM warmup

RData 408

1 Shot of es5

30 sec LDM warmup

RData 409

1 Shot of es5

30 sec LDM warmup

RData 410

es5 Micrometer moved to .00495

One shot

30 sec LDM warmup

RData 411

es5 Micrometer moved to .00400 (CERN photogrametry position)

One shot

30 sec LDM warmup

RData 412

Micrometer at .0000

1 shot es5

30 sec LDM warmup (Lights ON and Saturated Sensor)

RData 413

Micrometer at .0000

1 shot es5

30 sec LDM warmup

RData 414

Moving 4 notches on es5 to .0004

Es4 read out by mistake

RData 415

es5 at .0004

es5 readout only

RData 416

es5 at .0008

es5 readout only – forgot doghouse, saturated

RData 417

es5 at .0008

es5 readout only

RData 418

es5 at .0040 (moved 10*4 notches)

es5 readout only

RData 419

es5 at .0000

es5 readout only

RData 420

es7 original position at .2001

One Shot of es7

RData 421

Es7 moved to .2005

One Shot of es7 in Bi-Directional mode

Lights On w/ Doghouses

RData 422

Es7 moved to .2005

2 Clusters of es7 in Bi-Directional mode

Lights Off w/ Doghouses

90 sec interval, 30 sec LDM warmup

RData 423

Es7 moved to .2005

One Shot of es7 in Bi-Directional mode

Lights On w/ Doghouses

RData 424

Single Shot of es7 w/ micrometer at .2001

RData 425

Single Shot of es7 w/ micrometer at .2005

RData 426

Single Shot of es7 w/ micrometer at .2041

RData 427

Single Shot of es7 w/ micrometer at .2001

RData 428

Single Shot of es7 w/ micrometer at .2021

RData 429

Single Shot of es7 w/ micrometer at .2001

RData 430

Single Shot of es7 w/ micrometer at .1997

RData 431

Single Shot of es7 w/ micrometer at –40 notches from .2001

RData 432

Single Shot of es7 w/ micrometer at .2001 (original position)

RData 433

Single Shot of es7 w/ micrometer at –20 notches 2000

RData 434

Single Shot of es7 w/ micrometer at .2001 (original position)

RData 435

Single Shot of es7 w/ micrometer at .2001 (original position)

 

Interupting Shift Studies for a Break …

RData 436

Full SLM + Axial Line

13 Sensors Bi-Directional

6 Clusters 4:0 Laser ON: Laser Off

90 sec interval w/ 30 sec LDM warmup

Lights off (es7 micometer at .2001)

 

Returning to look at Shifts

RData 437

Single shot only, check to see if laser is moving, es7 micrometer at .2001

es3, es5, and es6 read out only

error in es3 – bad fit parameter

RData 438

Single shot only, check to see if laser is moving, es7 micrometer at .2001

es3, es5, and es6 read out only

RData 439

Single Shot (es 4-7 readout only)

es7 -> +8 clicks to .2009

es6

es5 -> +8 clicks to .0008

es4

RData 440

Single Shot (es 4-7 readout only)

es7 -> +20 clicks to .2021

es6

es5 -> +20 clicks to .0020

es4

RData 441

Single Shot (es 4-7 readout only)

es7 -> +40 clicks to .2041

es6

es5 -> +40 clicks to .0040

es4

RData 442

Single Shot (es 4-7 readout only)

es7 back at original position (.2001)

es6

es5 back at original position (.0000)

es4

 

Finished with Shift Studies

RData 443

3 Quick Bi-Directional Events

Full SLM + Axial Line

13 sensors

RData 444

3 Quick Bi-Directional Events

Full SLM + Axial Line

13 sensors

 

Looking at LDM warmup in LDM302 & LDM301

Single Shot  SLM + Axial Line

13 Sensors

RData 445

2 sec warmup time

RData 446

4 sec warmup time

RData 447

6 sec warmup time

RData 448

8 sec warmup time

RData 449

15 sec warmup time

RData 450

20 sec warmup time

RData 451

40 sec warmup time

RData 452

60 sec warmup time

RData 453

75 sec warmup time

RData 454

90 sec warmup time

RData 455

110 sec warmup time

RData 456

180 sec warmup time

 

End of June 2000 ISR Testing at CERN