Summary of June 2000 CERN ISR Test Data
Updated : 19 July 2000 RL
You can find the June 2000 ISR Test Data Description here.
Runs which have
been checked and/or corrected are marked in YELLOW.
|
|
Initial System Setup and Diagnostics
|
|
RData 210 |
LDM Power off, 6 sensors |
|
RData 211 |
LDM Power on, 6 sensors |
|
RData 212 |
error w/ DCOPS (2 trys) |
|
RData 323 |
new daq configuration, failed self test |
|
RData324 |
corrupt data w/ 14 sensors |
|
RData325 |
7 sensors, missing #0001 characters and sensor 1 data seems high |
|
RData 326 |
6 sensors, run looks ok |
|
RData 327 |
10 sensors, readout is ok, 10 cycles w/ 1 event per cycle (5 sec delay) |
|
|
HP Unit Disconnected – Manual LDM Operation |
|
RData 328 |
First trial w/ no Data Second Trial w/ HP Disconnected, manual LDM operation (left on) 10 sensors connected |
|
RData 329 |
Test w/ HP Disconnected |
|
RData 330 |
Testing 13 sensor readout 1 event, data looks like empty buffers (no bkground) |
|
RData 331 |
Testing 13 sensor readout 4 events, data looks like empty buffers |
|
RData 332 |
Testing 13 sensor readout 20 events, data looks like empty buffers |
|
RData 333 |
Test w/ HP Disconnected |
|
RData 334 |
Test w/ HP Disconnected |
|
RData 335 |
Test w/ HP Disconnected |
|
RData 336 |
Test w/ HP Disconnected |
|
RData 337 |
Test w/ HP Disconnected |
|
RData 338 |
Test w/ HP Disconnected |
|
RData 339 |
Axial Line Readout – No HP Data |
|
RData 340 |
Axial Line Readout
w/ HP Data |
|
RData 341 |
et0 readout only – looks like bkg only(?) |
|
RData 342 |
et0 readout only, 1 event |
|
|
Axial Line Surveyed, es7 micrometer moved from 2.90 -> 2.00 |
|
RData 343 |
20 cycles full Axial Line, No HP Data or Lasers On Data, abnormal termination after 16 cycles |
|
RData 344 |
Aborted as test |
|
|
Axial Line Complete - Axial Line Data Only w/ HP Unit Operational |
|
RData 345 |
20 clusters, 1 event, 20 sec delay, 5 sec LDM warmup |
|
RData 346 |
20 clusters, 1 event, 20 sec delay, 5 sec LDM warmup (small software correction) |
|
RData 347 |
1 clusters, 20 events 5 sec LDM warmup |
|
RData 348 |
Single Shot Axial Line |
|
RData 349 |
Single Shot Axial Line |
|
RData 350 |
Single Shot Axial Line |
|
RData 351 |
Single Shot Axial Line |
|
RData 352 |
Single Shot Axial Line |
|
RData 353 |
Axial LDM has been cold 3 hours, doghouses put on et 3&4, Lights off, 20 cycles of quick data on Axial Line only |
|
|
SLM Line Setup and Drilled Down |
|
RData 354 |
Full SLM line – 1 shot 13 sensors board reset in first event |
|
RData 355 |
Full SLM line + Axial line 10 clusters, 4 events per cluster, 60 sec interval abnormal termination, board reset in first event |
|
RData 356 |
Full SLM line + Axial line 2 x 1 shot Readout w/ Wux and CC Failed |
|
|
Splitting Lines into SLM and Axial LinesSingle Directional only unless noted |
|
RData 357 |
SLM Laser 303 – 5 events |
|
RData 358 |
2 sensor stability s6 and s1 (es10 & es5) 20 Quick Measurements to compare peak quality and resolution of a ‘poor’ sensor (s1) and a ‘good’ one (s6) No end run record, but data looks like it acquired ok |
|
RData 359 |
2 sensor stability s6 and s1 (es10 & es5) 10 Quick Measurements to compare peak quality and resolution of a ‘poor’ sensor (s1) and a ‘good’ one (s6) |
|
RData 360 |
2 sensor stability s6 and s1 (es10 & es5) 10 Quick Measurements to compare peak quality and resolution of a ‘poor’ sensor (s1) and a ‘good’ one (s6) |
|
RData 361 |
Full SLM w/ 5 events Abnormal termination in event 1 |
|
RData 362 |
Single Shot SLM 8 sensors readout |
|
RData 363 |
Single Shot SLM 8 sensors readout |
|
RData 364 |
2 Clusters SLM 1 Shot each |
|
RData 365 |
SLM Line 60 Clusters, 1 event per cluster, 30 sec LDM warmup, No end run record but readout of 60 clusters seems complete At least through #0002 of last event in last cluster |
|
RData 366 |
Aborted |
|
RData 370 |
Axial Line Single Shot Background Only (LDM was unplugged) 5 Sensors |
|
RData 371 |
Axial Line 13 sensors connected, 5 readout Cold laser, lights on, middle of afternoon |
|
RData 372 |
Axial Line ( 5 sensors connected, 5 readout only) 25 cycles looking at LDM warmup again - Lights on, Doghouses on Data initially has extra lines from DSP responses, but it appears limited to first event |
|
RData 373 |
Axial Line ( 5 sensors connected, 5 readout only) 20 cycles looking at LDM warmup again - Lights on, Doghouses on Data looks ok |
|
|
Final SLM Tape Configuration Applied |
|
RData 374 |
Axial Line 10 clusters w/ 3 events per cluster Computer hung after completion – Data looks ok on simple inspection |
|
RData 375 |
Axial Line running overnight 20 clusters w/ 4 events per cluster 60 sec spacing, 30 sec ldm warmup Run aborted early upon return the next morning. 15/20 Clusters completed |
|
|
Back to Single DCOPS String of 13 Sensors for SLM + Axial Lines |
|
RData 376 |
13 sensors Bi-Directional Laser 301 tied to Laser 303 4 Clusters 10:1 Laser ON:Laser OFF events 540 sec between clusters, 30 sec LDM warmup 84 events read |
|
|
Laser 301 now tied to Laser 303 AND Laser 302 Permanently |
|
RData 377 |
13 sensors Bi-Directional Laser 301 tied to Laser 303 AND Laser 302 Single Shot |
|
RData 378 |
13 sensors Bi-Directional 2 Clusters w/ 2 events each No time delay between clusters |
|
RData 379 |
13 sensors Bi-Directional Long Overnight Run 70 Clusters 5:1 Laser ON:Laser OFF events 150 sec between clusters, 30 sec LDM delay Killed next morning before lights turned on 38.5 clusters ok – Lights on at cluster 34; cluster 22 & 32 not corrected |
|
RData 380 |
Short Axial Run 20 clusters, 5 sensors only 60 sec delay, 30 sec LDM warmup |
|
RData 381 |
Short Test Run SLM + Axial Bi Directional 1 cluster w/ 1:1 laser on:laser off |
|
RData 382 |
13 sensors Bi-Directional 2 Clusters, 3:1 Laser ON:Laser OFF events 90 sec delay, 30 sec LDM Warmup |
|
|
Sensors Are Now Being Moved (RData 382 is original position) |
|
RData 383 |
et2 moved +1mm in +X direction (New position is 6mm) Bi-Dir 2 Clusters w/ 4 Laser On per cluster 90 sec delay, 30 sec LDM warmup Only 8 events taken (Single Shot Mode by mistake) |
|
RData 384 |
Repeating 383 Bi-Dir 2 Clusters w/ 4 Laser On per cluster 90 sec delay, 30 sec LDM warmup 16 events taken |
|
RData 385 |
et2 moved +1mm in +X direction (New position is 7mm) Bi-Dir 2 Clusters w/ 4 Laser On per cluster 90 sec delay, 30 sec LDM warmup Z Sensor on MAB is lost now from change in position System hung after 3 events |
|
RData 386 |
Repeating 385 Bi-Dir 2 Clusters w/ 4 Laser On per cluster 90 sec delay, 30 sec LDM warmup Z Sensor on MAB is lost now from change in position System hung after 3 events |
|
RData 387 |
et2 moved back to original position Bi-Dir 2 Clusters w/ 4 Laser On per cluster 90 sec delay, 30 sec LDM warmup |
|
RData 388 |
et2 moved -1mm in X direction Bi-Dir 2 Clusters w/ 4 Laser On per cluster 90 sec delay, 30 sec LDM warmup |
|
|
Motion Shift Studies on SLM Lineet2 moved back to 0 All Data taken with 2 Clusters, 4 events per Cluster w/ Laser ON 90 sec delay, 30 sec LDM warmup |
|
RData 389 |
es5 moved 100 um in +Z (4 notches) Previous position: .050 New Position: .054 |
|
RData 390 |
es5 moved 100 um in +Z (New Position: .058) |
|
RData 391 |
es5 moved 100 um in +Z |
|
RData 392 |
Back to .054 |
|
RData 393 |
Back to Original position (.050) |
|
RData 394 |
es5 moved 100 um in -Z (New Position: .046) 1 event only |
|
RData 395 |
Retry 394 w/ proper number clusters and events (described above) |
|
RData 396 |
Original micrometer position Aborted after 4 events by user (1 shot was intended) |
|
RData 397 |
Repeat 396 but w/ One Shot Only |
|
RData 398 |
Reading S5 only Moving 20 notches tp .5mm |
|
RData 399 |
Reading S5 only - One Shot at .055 |
|
RData 400 |
Reading S5 only - One Shot at .045 |
|
|
Ending Shift Studies |
|
RData 401 |
Long Run 18 Clusters 4 Events per Cluster 90 sec delay, 30 sec LDM warmup Lights Off w/ Doghouses on et 3&4 72 events read, but no end run record |
|
|
Looking at Shifts once Again |
|
RData 402 |
1 Shot of es5 (original position) Lights On / Dog Houses On ** We have had problems w/ micrometer (see log book pg 73) Micrometer original position is .00505 |
|
RData 403 |
1 Shot of es5 3 sec LDM warmup |
|
RData 404 |
1 Shot of es5 30 sec LDM warmup |
|
RData 405 |
1 Shot of es5 30 sec LDM warmup |
|
RData 406 |
1 Shot of es5 30 sec LDM warmup |
|
RData 407 |
1 Shot of es5 30 sec LDM warmup |
|
RData 408 |
1 Shot of es5 30 sec LDM warmup |
|
RData 409 |
1 Shot of es5 30 sec LDM warmup |
|
RData 410 |
es5 Micrometer moved to .00495 One shot 30 sec LDM warmup |
|
RData 411 |
es5 Micrometer moved to .00400 (CERN photogrametry position) One shot 30 sec LDM warmup |
|
RData 412 |
Micrometer at .0000 1 shot es5 30 sec LDM warmup (Lights ON and Saturated Sensor) |
|
RData 413 |
Micrometer at .0000 1 shot es5 30 sec LDM warmup |
|
RData 414 |
Moving 4 notches on es5 to .0004 Es4 read out by mistake |
|
RData 415 |
es5 at .0004 es5 readout only |
|
RData 416 |
es5 at .0008 es5 readout only – forgot doghouse, saturated |
|
RData 417 |
es5 at .0008 es5 readout only |
|
RData 418 |
es5 at .0040 (moved 10*4 notches) es5 readout only |
|
RData 419 |
es5 at .0000 es5 readout only |
|
RData 420 |
es7 original position at .2001 One Shot of es7 |
|
RData 421 |
Es7 moved to .2005 One Shot of es7 in Bi-Directional mode Lights On w/ Doghouses |
|
RData 422 |
Es7 moved to .2005 2 Clusters of es7 in Bi-Directional mode Lights Off w/ Doghouses 90 sec interval, 30 sec LDM warmup |
|
RData 423 |
Es7 moved to .2005 One Shot of es7 in Bi-Directional mode Lights On w/ Doghouses |
|
RData 424 |
Single Shot of es7 w/ micrometer at .2001 |
|
RData 425 |
Single Shot of es7 w/ micrometer at .2005 |
|
RData 426 |
Single Shot of es7 w/ micrometer at .2041 |
|
RData 427 |
Single Shot of es7 w/ micrometer at .2001 |
|
RData 428 |
Single Shot of es7 w/ micrometer at .2021 |
|
RData 429 |
Single Shot of es7 w/ micrometer at .2001 |
|
RData 430 |
Single Shot of es7 w/ micrometer at .1997 |
|
RData 431 |
Single Shot of es7 w/ micrometer at –40 notches from .2001 |
|
RData 432 |
Single Shot of es7 w/ micrometer at .2001 (original position) |
|
RData 433 |
Single Shot of es7 w/ micrometer at –20 notches 2000 |
|
RData 434 |
Single Shot of es7 w/ micrometer at .2001 (original position) |
|
RData 435 |
Single Shot of es7 w/ micrometer at .2001 (original position) |
|
|
Interupting Shift Studies for a Break … |
|
RData 436 |
Full SLM + Axial Line
13 Sensors Bi-Directional 6 Clusters 4:0 Laser ON: Laser Off 90 sec interval w/ 30 sec LDM warmup Lights off (es7 micometer at .2001) |
|
|
Returning to look at Shifts |
|
RData 437 |
Single shot only, check to see if laser is moving, es7 micrometer at .2001
es3, es5, and es6 read out only error in es3 – bad fit parameter |
|
RData 438 |
Single shot only, check to see if laser is moving, es7 micrometer at .2001
es3, es5, and es6 read out only |
|
RData 439 |
Single Shot (es 4-7 readout only) es7 -> +8 clicks to .2009 es6 es5 -> +8 clicks to .0008 es4 |
|
RData 440 |
Single Shot (es 4-7 readout only) es7 -> +20 clicks to .2021 es6 es5 -> +20 clicks to .0020 es4 |
|
RData 441 |
Single Shot (es 4-7 readout only) es7 -> +40 clicks to .2041 es6 es5 -> +40 clicks to .0040 es4 |
|
RData 442 |
Single Shot (es 4-7 readout only) es7 back at original position (.2001) es6 es5 back at original position (.0000) es4 |
|
|
Finished with Shift Studies |
|
RData 443 |
3 Quick Bi-Directional Events Full SLM + Axial Line 13 sensors |
|
RData 444 |
3 Quick Bi-Directional Events Full SLM + Axial Line 13 sensors |
|
|
Looking at LDM warmup in LDM302 & LDM301Single Shot SLM + Axial Line 13 Sensors |
|
RData 445 |
2 sec warmup time |
|
RData 446 |
4 sec warmup time |
|
RData 447 |
6 sec warmup time |
|
RData 448 |
8 sec warmup time |
|
RData 449 |
15 sec warmup time |
|
RData 450 |
20 sec warmup time |
|
RData 451 |
40 sec warmup time |
|
RData 452 |
60 sec warmup time |
|
RData 453 |
75 sec warmup time |
|
RData 454 |
90 sec warmup time |
|
RData 455 |
110 sec warmup time |
|
RData 456 |
180 sec warmup time |
|
|
End of June 2000 ISR Testing at CERN |