adc monitoring in Selex

16-Apr-1997
2 fastbus crates were used in Selex experiment to degitize and read out data from 3 led glass photon detectors and the neutron calorimeter. Those detectors were continuously (every 2-4 hours) calibrated using red and green LED and blue laser. The monitoring program (mon) worked on fastbus CPU - fscc and transfered the calibration data to host computer. Several camac units were used to make LED and Laser trigger, proper delay etc. In off line Soap analysis those data are used in compressed form of OCS tables to setup PM gains, to make a tables of noisy and dead counters.
vmatveev@fnal.gov